Features

  • Traceability according requirements of ISO/TS 16949 or ISO 9001:2008
  • Calibration maschines and accredited according ISO 17025 with scope from 2N to 250kN.
  • Highest precision and low uncertainty (CMC better 0,02%)

Description

Verification and calibration of force measurement systems (force gauge, load cell with amplifier and indicator) traceable to PTB/DKD/ILAC standards.

The directive DAkkS-DKD-R 3-3 describes calibration procedures for force measuring systems. The reduced scope according procedure "B" calibrates the object in one position with 2 measuring rows with increasing and decreasing force levels at 5 points. This procedure can apply if "a-priori" knowledge is available to calculate the uncertainty accordingly.

The calibration for tensile and compression force (pull and push direction) is performed on an automated calibration stand with direct staggered mass application for forces up to 1000N. The lowest resolution is 0,2N for force gauges with 2N measuring range. Force gauges with nominal measuring range higher 1kN are calibrated on a stand with transfer standards up to 250 kN. Please consult the below table in order to check which measuring ranges of the calibration stands are covered by the accreditation in relation to the nominal force range of your gauge.

Models and Prices

Select Part no. Direction Range F(nom) CMC (accredited) Availability Price €
CAL-210A5B Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug/Druck 1-5 N - 295,00
CAL-210A5D Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Druck 1-5 N - 215,00
CAL-210A5Z Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug 1-5 N - 215,00
CAL-210B5B Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug/Druck 10-50 N 0,02 % 315,00
CAL-210B5D Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Druck 10-50 N 0,02 % 235,00
CAL-210B5Z Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug 10-50 N 0,02 % 235,00
CAL-210K1B Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug/Druck 100-1000 N 0,02 % 315,00
CAL-210K1D Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Druck 100-1000 N 0,02 % 235,00
CAL-210K1Z Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug 100-1000 N 0,02 % 235,00
CAL-210K5B Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug/Druck 25-5000 N 0,1 % 315,00
CAL-210K5D Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Druck 25-5000 N 0,1 % 235,00
CAL-210K5Z Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug 25-5000 N 0,1 % 235,00
CAL-210M2B Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug/Druck 2500-25000 N 0,1 % 315,00
CAL-210M2D Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Druck 2500-25000 N 0,1 % 235,00
CAL-210M2Z Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug 2500-25000 N 0,1 % 235,00
CAL-210T2B Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug/Druck 30000-250000 N 0,1 % 315,00
CAL-210T2D Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Druck 30000-250000 N 0,1 % 235,00
CAL-210T2Z Calibration Certificate acc. DAkkS-DKD-R 3-3 (B) Zug 30000-250000 N 0,1 % 235,00

Didn't find the right thing?

Our Solution Finder might help you!